Climate-Ready External-Beam Radiotherapy in Viet Nam: Treatment Interruptions, Quality Assurance, and Safe Resumption

IMPLEMENTATION INSIGHTSCLIMATE-RESILIENT HEALTHCARE

11/5/202624 min read

Climate-Ready External-Beam Radiotherapy in Viet Nam: Treatment Interruptions, Quality Assurance, and Safe Resumption

IMPLEMENTATION INSIGHTS · CLIMATE-RESILIENT HEALTHCARE

Last reviewed: November 5, 2026

For someone receiving radiotherapy for cancer, an unexpected interruption can mean more than a missed hospital appointment. It may affect a carefully planned course of treatment in which the timing, delivery, and recording of each radiation session are clinically important.

Consider a patient receiving external-beam radiotherapy over several weeks. Heavy rainfall disrupts hospital operations, leaving the treatment unit unavailable and several scheduled sessions missed. Once the equipment has been repaired, the engineering team confirms that the technical fault has been resolved. The radiation oncology department must then determine whether the machine is ready for clinical use and how the patient's treatment should continue.

For some patients, a short interruption may have limited clinical consequences. For others, extending the overall treatment time could affect the intended therapeutic outcome. The appropriate response cannot be determined simply by adding missed appointments to the end of the schedule.

Radiotherapy depends on much more than a functioning machine. Its safety and effectiveness rely on accurate treatment planning, radiation delivery, patient positioning, reliable treatment records, quality assurance, and the professional decisions that connect these activities.

For healthcare organizations in Viet Nam, climate readiness therefore involves both technical recovery and the continuity of individual treatment courses. It requires understanding how a disruption may affect ongoing care, which information must remain reliable, and what clinical and technical conditions need to be established before treatment resumes.

The objective is to minimize avoidable interruptions while ensuring that patients continue receiving safe, accurate, and clinically appropriate treatment.

Why Radiotherapy Requires a Distinct Approach to Climate Resilience

External-beam radiotherapy delivers radiation from equipment located outside the patient's body to treat a defined target. Depending on the clinical indication and available technology, treatment may involve photons, electrons, or other external radiation modalities.

Many radiotherapy centers use medical linear accelerators to deliver photon or electron treatments. These machines operate within a specialized environment involving clinical assessment, treatment planning, dose calculation, imaging, patient positioning, verification, and radiation delivery.

For many patients, the prescribed radiation dose is divided into multiple treatment sessions known as fractions. The number of fractions, treatment schedule, technique, and overall duration depend on the disease, treatment intent, and individual clinical circumstances.

These arrangements have therapeutic significance. The biological effects of radiotherapy depend partly on how radiation is delivered over time, and treatment interruptions may have different consequences depending on the cancer and prescribed regimen.

Radiotherapy also relies on interconnected technical and information systems. A treatment unit may depend on imaging equipment, patient positioning components, treatment planning software, oncology information systems, and accurate records of radiation already delivered.

A problem affecting one essential component may prevent treatment even when the hospital remains physically open and other services continue operating.

This makes radiotherapy particularly sensitive to disruptions that affect not only equipment but also the sequence and integrity of treatment. Restoring machine availability is important, but it does not necessarily resolve the implications of an interruption for the patient.

A meaningful approach to climate resilience must therefore consider both the capability to deliver radiation accurately and the continuity of the patient's intended course of care.

How Severe Weather Can Disrupt Radiotherapy

Climate-related hazards can affect radiotherapy through several pathways.

Heavy rainfall and flooding may restrict access to treatment facilities, damage technical areas, or interrupt supporting infrastructure. Severe storms can affect electricity and communications, while prolonged high temperatures may create challenges for maintaining the operating conditions required by sensitive equipment.

Some disruptions affect the treatment machine directly. Others interfere with supporting systems or prevent patients and healthcare personnel from reaching the treatment center.

A department may have functioning radiation delivery equipment but temporarily lose access to an information system required for treatment verification. Another center may retain operational technical systems while flooding prevents patients from attending scheduled appointments.

These situations create different risks and require different responses.

Equipment faults may need engineering and medical physics assessment. Problems involving treatment information raise concerns about whether the correct plan and delivery history remain available and reliable. Difficulties with patient access may require clinical prioritization, communication, and appropriate coordination.

Not every interruption occurring during severe weather is necessarily caused by the weather itself, and an individual event should not automatically be attributed to climate change. Equipment defects, software problems, maintenance activities, staffing shortages, and other operational failures may create similar challenges.

Understanding the cause remains important for prevention and improvement. However, the immediate clinical concern is whether treatment has been interrupted and what is necessary for safe continuation.

The timing and nature of the interruption also matter. A delay before the first treatment fraction is different from a gap during an established course. Similarly, a canceled appointment is not equivalent to a machine fault occurring after radiation delivery has begun.

These differences must be understood before decisions are made about subsequent treatment.

The Clinical Significance of Missed Treatment Fractions

An important principle in radiotherapy continuity is that interruptions do not affect every patient in the same way.

Their potential consequences depend on factors including cancer type, treatment intent, prescribed fractionation, timing and duration of the interruption, and the patient's clinical condition.

For some cancers treated with curative intent, prolonging the overall treatment time may reduce the probability of achieving the intended tumor control. Other treatment regimens may be less sensitive to a particular interruption or may require different clinical considerations.

Palliative radiotherapy may also involve objectives and schedules that differ from treatment intended to achieve long-term disease control.

It would therefore be misleading to suggest that every missed fraction produces the same amount of harm or requires an identical corrective action.

The Royal College of Radiologists' Timely Delivery of Radical Radiotherapy: Guidelines for the Management of Unscheduled Treatment Interruptions, fourth edition, published in 2019, provides an established international reference on this subject.

The guidance considers the effects of unscheduled interruptions according to tumor type, treatment intent, and the circumstances of the delay. It discusses patient categorization and approaches to managing the consequences of unavoidable treatment prolongation.

The fourth edition remains a useful professional reference while an updated edition is under development. However, its recommendations should not be interpreted as a mandatory Vietnamese clinical protocol or as authorization to change an individual patient's prescribed treatment without appropriate assessment and approval.

The essential principle is that interruption management must remain patient-specific.

A department may need to review several patients affected by the same operational event, but the appropriate response for one patient may differ substantially from that for another.

A Missed Fraction Is Different From Partially Delivered Treatment

An interruption before radiation delivery begins raises different concerns from an unexpected machine shutdown after part of the prescribed radiation has already been delivered.

When a patient misses a scheduled fraction, the treatment record should accurately indicate that the fraction was not delivered.

A more complicated situation arises when equipment stops during treatment. Some radiation may already have been administered, and the clinical team needs reliable information about what occurred.

A scheduled appointment does not establish that treatment was delivered. Likewise, a system indicating an incomplete session may not, by itself, provide sufficient information to determine the radiation already administered.

This becomes particularly important when equipment faults are accompanied by incomplete records, communication failures, or inconsistent information between connected systems.

The American Association of Physicists in Medicine (AAPM) addressed these issues in AAPM Task Group Report 314: Fault Recovery in External Beam Radiation Therapy, published online in November 2024 and in the January 2025 issue of Medical Physics.

The report examines recovery from fault conditions involving medical linear accelerators and associated systems, with particular attention to equipment faults, interrupted radiation delivery, and the information necessary for safe treatment continuation.

An important principle is that the status of an interrupted fraction must be established through appropriate evidence rather than assumptions.

Where the available information does not adequately establish what was delivered, qualified personnel must assess the technical and dosimetric implications before further radiation is administered.

Automatically repeating a fraction or delivering a presumed remaining dose without the necessary professional evaluation could introduce additional risks.

These are specialized clinical and medical physics decisions. They cannot be treated as routine scheduling adjustments.

Treatment Continuation Requires Individual Clinical Judgment

Following a disruption, radiotherapy departments may face considerable pressure to restore treatment schedules.

Patients may be anxious about missed appointments, while clinical teams may be concerned about unnecessary prolongation of treatment. These concerns are legitimate, but extending operating hours, adding appointments, or transferring patients to another machine does not automatically resolve the clinical implications of an interruption.

The responsible radiation oncologist may need to review the patient's treatment history, prescribed course, treatment intent, interruption duration, and current clinical condition.

A qualified medical physicist may also need to assess technical or dosimetric implications, particularly where treatment delivery was interrupted or changes to the treatment arrangement are being considered.

For some patients, continuation of the original treatment plan may remain appropriate without modification. Others may require additional clinical review or consideration of a different arrangement.

There is no universal recovery method suitable for every patient.

The existence of missed sessions does not automatically justify increasing the dose per fraction, adding fractions, changing treatment days, or modifying the prescribed technique.

Such changes may affect the balance between tumor control and radiation exposure to healthy tissues. Any clinically indicated modification therefore requires appropriate assessment, planning, verification, and authorization.

Healthcare organizations need to distinguish the operational objective of minimizing delays from the professional responsibility to preserve treatment safety and intended effectiveness.

Restoring Equipment Function Is Not the Same as Restoring Treatment Readiness

A radiotherapy machine may restart successfully after a fault while still requiring assessment before returning to clinical use.

This distinction is particularly important following hardware failure, software problems, water exposure, major repairs, or other events that may affect equipment performance.

The appropriate level of assessment depends on the circumstances.

A brief interruption that does not compromise relevant operating conditions may require a different response from a major component replacement or suspected damage involving radiation delivery, imaging, or patient positioning.

Not every interruption requires complete recommissioning of the treatment system. However, successful equipment startup or confirmation that a repair has been completed does not establish that every intended treatment can safely resume.

AAPM Task Group Report 314 emphasizes identifying the fault, understanding the corrective action, assessing possible consequences, and completing appropriate quality assurance before affected equipment returns to clinical service.

The report also recognizes the importance of manufacturer information and technical service expertise, while maintaining the need for professional judgment by qualified medical physics personnel.

For healthcare organizations, technical repair and clinical release are connected but distinct responsibilities.

A service engineer may confirm that an equipment defect has been corrected. Medical physics personnel must assess the relevant technical and dosimetric conditions, while clinical leadership determines whether appropriate patient treatment can resume within the facility's authorized arrangements.

These decisions should be supported by documented evidence rather than an assumption that restored mechanical operation establishes treatment safety.

Quality Assurance Must Address the Actual Disruption

Quality assurance in radiotherapy includes activities supporting accurate treatment planning, equipment performance, radiation delivery, imaging, and patient positioning.

It should not be understood as a single test capable of establishing the reliability of every part of a radiotherapy system under all circumstances.

Different faults require different forms of assessment.

A problem affecting beam delivery may require different verification from one involving imaging equipment, the treatment couch, software interfaces, or communication between systems.

Similarly, an interruption without physical damage may require a different evaluation from an event involving water intrusion or extensive technical repair.

Depending on the circumstances, appropriately qualified personnel may need to determine whether previously established performance characteristics remain valid and whether additional testing, calibration, recommissioning, or other technical action is necessary.

The required activities should reflect the identified problem and its potential consequences for the intended treatment.

A routine quality assurance record may provide useful evidence, but it does not automatically establish that a newly identified fault has been resolved.

The essential question is whether the available evidence adequately addresses the concerns arising from the disruption and supports the proposed clinical use.

Preserving Treatment Plans and Delivery Records

External-beam radiotherapy depends on accurate information throughout the treatment pathway.

Clinical assessment, simulation, target definition, treatment planning, dose calculation, plan approval, patient positioning, and radiation delivery are connected activities. An error or inconsistency affecting one part of this process may influence the treatment subsequently delivered.

During an environmental disruption, the integrity of treatment information becomes particularly important.

For example, a department may retain access to its treatment planning system while temporarily losing access to the oncology information system used to manage treatment delivery records.

Another center may restore its electronic systems after an interruption but need to determine whether the correct treatment information has been preserved.

The concern is not simply whether electronic files remain accessible. The organization must have sufficient confidence that the correct patient, approved treatment plan, treatment history, and necessary verification information remain accurately associated with one another.

A treatment plan stored in a computer system does not automatically establish that it is the latest approved version. Similarly, a completed appointment entry does not necessarily prove that the prescribed fraction was delivered correctly.

These distinctions matter because inaccurate treatment information may affect subsequent radiation delivery.

A partially delivered fraction may require additional technical assessment. An unresolved discrepancy between an approved plan and the information available to the treatment unit may prevent safe continuation until the concern has been addressed.

Radiotherapy therefore introduces technical and dosimetric considerations beyond ordinary electronic health record continuity.

Restoring information technology services should not automatically be interpreted as confirmation that the radiotherapy treatment pathway is ready to resume.

The necessary clinical and technical verification must be completed according to the facility's approved procedures.

When a Patient's Condition Changes During an Interruption

A prolonged interruption may create an additional question: does the original treatment arrangement remain appropriate for the patient?

During a treatment gap, a person's clinical condition or anatomy may change.

Depending on the disease and treatment technique, changes in body weight, tumor dimensions, surrounding tissue anatomy, or other relevant characteristics may affect assumptions used during treatment planning.

This does not mean that every interruption requires new simulation imaging or a revised treatment plan.

The need for reassessment depends on the individual patient, treatment technique, duration of the interruption, and available clinical information.

The responsible radiation oncologist may determine whether additional clinical evaluation is necessary before treatment continues.

Where changes could affect treatment geometry or dose distribution, medical physics and treatment planning expertise may also be required.

The decision should reflect the actual circumstances rather than a fixed rule applied to every patient.

Safe resumption depends not only on the treatment equipment functioning correctly, but also on whether the planned treatment remains appropriate for the person receiving it.

Coordinating Treatment Between Radiotherapy Centers

When a radiotherapy center cannot restore services promptly, transferring treatment to another authorized facility may be considered.

This may be particularly important when prolonged interruption could have significant clinical consequences.

However, transferring radiotherapy involves considerably more than arranging an appointment at another hospital.

The receiving center needs sufficient information about the patient's diagnosis, treatment intent, approved plan, radiation already delivered, and current clinical condition.

It must also determine whether it has the necessary equipment, professional capabilities, and authorization to deliver the intended treatment safely.

Radiotherapy equipment and treatment planning systems are not automatically interchangeable.

A plan developed for one treatment unit may not be directly transferable to another, especially when equipment characteristics, available techniques, or software systems differ.

The receiving center may need additional planning, verification, or other professional activities before treatment can continue.

These decisions require appropriate coordination among radiation oncologists, medical physicists, radiation therapists, and other relevant personnel.

The availability of an alternative machine does not establish that a patient can immediately continue the original plan.

Clinical responsibilities, treatment information, and relevant documentation must be managed appropriately. In Viet Nam, differences in facility capabilities and service availability also need to be considered when assessing whether an alternative treatment arrangement is feasible.

Patient privacy and confidentiality remain important throughout the process.

The Law on Personal Data Protection No. 91/2025/QH15 and Government Decree No. 356/2025/ND-CP, both effective from January 1, 2026, provide an important legal context for processing and sharing personal information.

Urgent treatment coordination does not automatically permit unrestricted disclosure or transfer of identifiable patient information.

Any arrangement must comply with applicable legal requirements, professional responsibilities, and organizational safeguards.

The objective is to support appropriate cancer care, not simply to move patients from one waiting list to another.

Communication With Patients Is Part of Treatment Continuity

Unexpected interruptions to radiotherapy can create considerable uncertainty for patients and families.

Patients may worry that treatment will become less effective, that missed fractions must be repeated, or that their original treatment plan is no longer suitable.

These concerns deserve clear and respectful communication.

At the same time, healthcare organizations should avoid generalized assurances that every interruption is harmless or that treatment can always resume without clinical consequences.

Communication should reflect the information actually available.

When treatment is temporarily unavailable, patients should receive reliable updates about service arrangements and how further information will be provided.

The responsible clinical team should explain any patient-specific implications that have been established through professional assessment.

If additional review is required before treatment continues, patients should understand that the review is intended to support safe and appropriate care.

Patients should not be encouraged to independently change their prescribed schedules or seek unverified replacement treatment arrangements.

Environmental disruption may also create practical difficulties involving transportation, employment, caregiving responsibilities, and other commitments.

These circumstances can affect a patient's ability to attend revised appointments.

Where suitable services exist, patient navigation and support personnel may help identify practical barriers and facilitate communication with the treating team.

Such support should complement clinical decision-making rather than replace it.

Maintaining reliable communication and appropriate follow-up can help preserve continuity of care even when treatment cannot immediately resume.

What Recent International Evidence Adds

Recent research provides a more detailed understanding of how disasters affect radiotherapy services and why recovery should be considered across the entire treatment pathway.

In September 2025, Müller-Polyzou and Reuter-Oppermann published Radiotherapy Continuity for Cancer Treatment: Lessons Learned From Natural Disasters in PLOS One.

The researchers conducted a structured literature review and an international survey involving radiotherapy professionals. Their work identified published experiences involving hurricanes, floods, and earthquakes, together with operational challenges affecting the delivery of radiotherapy.

The study developed a conceptual framework for understanding vulnerabilities and considering measures that may strengthen treatment continuity.

Its findings reinforce the importance of coordination between clinical, technical, organizational, and patient-related activities.

However, the evidence was drawn from a limited body of disaster-specific publications and professional survey responses. It should not be interpreted as proof that every proposed measure improves clinical outcomes or will work equally well across healthcare settings.

In September 2026, Han-Oh and colleagues published Building Resilient Radiotherapy Systems: Lessons and Recommendations for Natural Disaster Risk Mitigation in Seminars in Radiation Oncology.

This narrative review synthesized published literature, incident reports, and operational guidance concerning natural disasters and related technological disruptions.

The authors organized proposed strategies across eight areas, including emergency governance, infrastructure, utilities and technical continuity, information systems, clinical care, treatment interruption management, workforce readiness, and patient support.

The review highlights why radiotherapy resilience should be understood as an interconnected system rather than exclusively as an equipment protection problem.

Another relevant publication from September 2026 is Emergency Preparedness for External Beam Radiotherapy Machines: Treatment Continuity and Interruption Recovery in Radiation Oncology, by Li, Anderson, and Rong.

This review specifically examines preparedness and recovery for external-beam radiotherapy equipment within the wider treatment workflow. It addresses issues such as patient-centered prioritization, the capabilities of available treatment systems, contingency arrangements for imaging and treatment information, partially delivered fractions, and appropriate return-to-treatment assessment.

Its focus is particularly relevant because a technically functioning machine does not necessarily mean the complete treatment pathway has been restored.

In the same journal, Ford's Resilience to Disaster: System Design and Future Directions discusses disaster preparedness using a framework based on healthcare structure, processes, and outcomes.

Taken together, these publications support a more comprehensive approach to radiotherapy preparedness and recovery.

However, they are primarily reviews, frameworks, and professional recommendations. They should not be presented as universally validated interventions or as guarantees of improved cancer outcomes.

For Viet Nam, their value lies in helping healthcare organizations identify relevant risks, understand professional responsibilities, and consider approaches suited to their local circumstances.

They do not justify applying identical interruption-management arrangements to every cancer diagnosis, treatment regimen, or radiotherapy center.

International Radiation Safety Guidance and Professional Responsibilities

The International Atomic Energy Agency (IAEA) provides an established international framework for radiation protection and safety in medical uses of ionizing radiation.

An important reference is Radiation Protection and Safety in Medical Uses of Ionizing Radiation, IAEA Specific Safety Guide No. SSG-46, published in 2018 and jointly sponsored by IAEA, the International Labour Organization, the Pan American Health Organization, and WHO.

The guide addresses medical, occupational, and public exposure to ionizing radiation.

It provides recommendations concerning radiation protection, optimization of safety, quality assurance, competent personnel, and the responsibilities of organizations using radiation in healthcare.

For radiotherapy, multidisciplinary responsibilities are particularly important.

Radiation oncologists are responsible for clinical treatment decisions. Qualified medical physicists provide specialist expertise concerning radiation physics, dosimetry, equipment performance, and relevant quality assurance. Radiation therapists carry out treatment-related activities within their authorized professional roles and approved procedures.

Healthcare facilities and their management have responsibilities for the organizational arrangements, resources, and safety systems necessary to support these activities.

These roles are complementary.

Technical personnel cannot independently determine every clinical consequence of a treatment interruption. Likewise, a clinical decision that treatment is urgently needed does not remove the requirement for appropriate equipment and medical physics assessment.

IAEA guidance provides a useful international reference, but its recommendations do not automatically become binding Vietnamese legal requirements merely because they have been published.

Their application must be considered alongside national legislation, relevant licensing conditions, professional responsibilities, and the healthcare organization's approved procedures.

Viet Nam's Updated Atomic Energy and Radiation Safety Framework

Viet Nam introduced significant changes to its atomic energy and radiation safety framework in 2026.

The Law on Atomic Energy No. 94/2025/QH15, dated June 27, 2025, took effect on January 1, 2026.

It establishes a statutory framework covering atomic energy applications, radiation protection and safety, relevant licensing and regulatory oversight, and incident response, among other matters.

For healthcare organizations operating radiotherapy equipment, the law provides an important legal foundation.

Government Decree No. 332/2025/ND-CP, dated December 18, 2025, also took effect on January 1, 2026.

It provides detailed provisions and implementation measures concerning radiation and nuclear safety, security, declaration, licensing, inspection, and relevant incident response activities within its scope.

Another important instrument is Ministry of Science and Technology Circular No. 59/2025/TT-BKHCN, dated December 31, 2025, effective January 1, 2026.

The circular regulates radiation safety and response to radiation and nuclear incidents.

These instruments are relevant when healthcare organizations consider the operation, maintenance, repair, modification, and safe use of radiotherapy equipment.

However, their requirements must be interpreted according to the radiation practice, equipment, authorization, licensing conditions, and circumstances involved.

An operational interruption does not automatically constitute a reportable radiation incident under Vietnamese law.

For example, an equipment shutdown that prevents treatment may be an operational failure without necessarily meeting the legal criteria for a radiation incident.

Conversely, an event involving suspected unintended irradiation, loss of relevant safety controls, or other circumstances falling within the applicable legal definition may trigger specific investigation, notification, or response obligations.

The distinction must be determined by responsible and competent parties under the relevant legal provisions.

It should not be inferred solely because a radiotherapy machine stopped during severe weather.

Healthcare organizations must also comply with the Law on Medical Examination and Treatment No. 15/2023/QH15, effective January 1, 2024, as amended, and its relevant implementing regulations.

Authorization to operate radiation equipment and authorization to provide clinical radiotherapy are related but distinct matters.

Compliance with one does not automatically establish compliance with the other.

Viet Nam's Radiotherapy Technical Guidance in 2026

The Ministry of Health issued important updates to radiotherapy technical guidance in June 2026.

Decision No. 1585/QD-BYT, dated June 3, 2026, promulgated Technical Procedure Guidelines for Radiotherapy, Volume 1, containing 34 technical procedures.

The decision took effect on its date of issuance and provides a current domestic technical reference for healthcare facilities within its scope.

It also repealed specifically identified procedures from the earlier oncology technical guidance issued under Decision No. 3338/QD-BYT, dated September 9, 2013.

Consequently, the earlier guidance should not be cited as though all its provisions remained unchanged after June 3, 2026.

On the same date, the Ministry issued Decision No. 1586/QD-BYT, promulgating Technical Procedure Guidelines for Radiotherapy, Volume 2, containing eight technical procedures.

An important subsequent amendment concerned the date from which Volume 2 would apply.

On July 1, 2026, the Ministry issued Decision No. 1986/QD-BYT, amending the implementation or application dates of several technical guideline decisions.

For radiotherapy Volume 2, the decision changed the application date to January 1, 2028.

Therefore, as of November 5, 2026, the two radiotherapy guideline volumes have different application statuses.

Volume 1 has been issued and is effective, while the application of Volume 2 is scheduled for January 1, 2028 under the amended provisions.

This distinction is important for accurate regulatory interpretation. The two volumes should not be presented as though both were already applicable in November 2026.

The guidance also covers radiotherapy procedures more broadly and should not be interpreted as a single comprehensive protocol for managing climate-related treatment interruptions.

Equally, inclusion of a technical procedure in Ministry of Health guidance does not automatically authorize every healthcare facility to perform it.

The actual provision of radiotherapy depends on applicable legal requirements, authorized services, professional competence, suitable equipment, and other relevant conditions.

These domestic references are important for healthcare organizations reviewing their clinical and technical arrangements, but they do not replace obligations arising under radiation protection and safety legislation.

Safe Resumption Requires Coordinated Professional Decisions

When a radiotherapy service is interrupted, several professional groups may need to contribute to decisions about resumption.

Engineering personnel may investigate and correct equipment faults. Medical physics personnel may determine the technical and dosimetric verification required before affected equipment returns to clinical use. Radiation oncologists may assess the implications of missed or partially delivered treatment and determine the appropriate clinical course for individual patients.

Radiation therapists may provide information about treatment delivery, positioning, recorded interruptions, and operational concerns. Information technology personnel may be involved when treatment planning systems, communication interfaces, or clinical records have been affected.

Healthcare leaders may need to coordinate resources, service availability, and organizational decisions.

These responsibilities are connected, but they should not be confused.

A manufacturer's service engineer may confirm that equipment repair has been completed, but this does not automatically establish that all planned radiotherapy treatments can resume.

Similarly, a physician's determination that continuing treatment is clinically important does not remove the need for appropriate equipment and medical physics quality assurance.

The organization must connect these decisions through established professional and governance arrangements.

A useful distinction is between technical readiness, treatment information integrity, and patient-specific clinical readiness.

Each may be established at a different time.

A machine may be technically suitable for treatment while an individual patient's course still requires clinical review. Another patient may have a clinically appropriate plan but remain unable to continue until necessary equipment verification has been completed.

Recognizing these differences helps avoid unsupported assumptions and supports decisions grounded in appropriate evidence.

The objective is not to create unnecessary delays. It is to ensure that treatment resumes with the required professional assessment and authorization.

Measuring Treatment Continuity and Learning From Disruption

Radiotherapy departments may monitor machine availability, canceled appointments, treatment interruptions, and the time required to restore services.

These measures provide useful information about operations, but they do not fully describe the clinical consequences of a disruption.

Two departments may report the same number of missed sessions, yet the implications may differ substantially according to the patients involved, their prescribed regimens, and the timing of the interruptions.

Similarly, a machine may return to service quickly while some patients still require clinical reassessment or treatment plan review.

A more meaningful evaluation therefore considers both operational continuity and the safety of treatment resumption.

Depending on the facility and available information, relevant measures may include the number of affected treatment courses, interruption duration, unresolved treatment records, equipment faults requiring investigation, and completion of appropriate quality assurance before clinical use.

Organizations may also examine whether patient-specific reviews were completed where necessary and whether appropriate coordination occurred.

These indicators require careful interpretation.

An increase in documented equipment problems after improvements in incident reporting does not necessarily mean equipment safety has deteriorated.

Similarly, the absence of reported treatment errors does not establish that every interruption was managed appropriately.

The purpose of monitoring is to support learning rather than merely demonstrate that equipment has returned to operation.

A disruption may reveal weaknesses that were not obvious during routine care. One department may discover uncertainty about responsibilities for reviewing partially delivered fractions. Another may identify gaps in escalation arrangements after equipment faults. A third may find that transferring treatment information between centers is more difficult than anticipated.

These findings can inform training, documentation, communication, and organizational improvement.

Some improvements may involve clearer professional responsibilities. Others may require specialized technical assessment, additional resources, or changes to established procedures.

The appropriate response depends on the problem identified and must remain consistent with patient safety and radiation protection requirements.

The value of resilience monitoring lies in understanding whether the organization can recognize significant risks, respond appropriately, and support safe treatment continuity.

Environmental Sustainability and Responsible Radiotherapy Operations

Radiotherapy services also have environmental and resource implications.

Treatment equipment and supporting infrastructure may consume substantial electricity and depend on specialized environmental controls.

Appropriate equipment maintenance, scheduling, and resource management may offer opportunities to improve operational efficiency and reduce unnecessary waste.

However, environmental objectives must remain compatible with treatment accuracy, radiation protection, and clinical care.

Energy-saving measures should not interfere with essential quality assurance, required equipment operating conditions, or patient safety systems.

Likewise, consolidating appointments or reducing operating hours should not justify changes to prescribed radiation treatment without appropriate clinical assessment.

A measure that appears environmentally beneficial may be unsuitable if it increases treatment delays or introduces additional safety risks.

For Climate & Health Vietnam, environmental sustainability is therefore a complementary consideration rather than the primary purpose of radiotherapy interruption management.

The immediate priority is to protect patients and preserve safe, appropriate cancer treatment.

What Climate & Health Vietnam Can Contribute

Climate-ready external-beam radiotherapy is an important subject for Climate & Health Vietnam (CHVN) because it connects environmental disruption with specialist clinical care, healthcare quality, patient safety, and organizational governance.

The topic illustrates why climate resilience cannot be evaluated solely through the condition of physical infrastructure or equipment availability.

A radiotherapy center may have a functioning treatment unit while individual treatment courses still require professional review. Another center may have clinically appropriate treatment plans but remain unable to resume care until equipment performance and treatment information have been adequately verified.

Understanding these situations requires collaboration among radiation oncologists, medical physicists, radiation therapists, biomedical engineers, healthcare quality professionals, hospital leaders, and other relevant specialists.

CHVN can contribute by synthesizing credible international evidence, explaining the distinction between technical recovery and clinical treatment continuity, and examining how these principles relate to Viet Nam's professional and regulatory context.

The platform can also support professional learning and discussion about the responsibilities involved in safe service resumption.

Such activities must respect the specialized nature of radiotherapy.

International case studies and professional guidance may provide valuable lessons, but they should not be presented as universal solutions or mandatory Vietnamese treatment protocols.

CHVN's role remains focused on evidence synthesis, professional education, knowledge sharing, and responsible collaboration.

It does not provide radiotherapy treatment, medical physics services, radiation safety inspections, equipment certification, or regulatory approval.

Nor does CHVN claim authority to prescribe or modify radiation treatment, calculate compensatory doses, approve treatment plans, authorize the resumption of individual treatment courses, or determine whether a radiotherapy facility complies with applicable requirements.

Its contribution is to make relevant knowledge accessible and encourage informed discussion among the professionals and organizations responsible for safe cancer care.

Safe Recovery Means Protecting the Treatment, Not Just Restoring the Machine

A climate-related disruption can affect radiotherapy in ways that are not immediately visible when a treatment unit returns to operation.

A machine may restart successfully while technical verification remains incomplete. A fraction may have been interrupted without sufficient information about radiation already delivered. A treatment plan may remain available while the patient's clinical circumstances require reassessment.

A center may also have appointment capacity again while some patients still need individualized decisions about treatment continuation.

These situations demonstrate why restoring radiotherapy services is more complex than restoring equipment.

For healthcare organizations in Viet Nam, responsible recovery requires understanding the interruption, identifying the affected clinical and technical activities, and determining whether sufficient evidence supports safe continuation.

Not every missed fraction requires the same response, not every operational interruption constitutes a radiation incident, and not every completed repair establishes that all treatment courses can immediately resume.

Decisions must reflect patients' clinical needs, treatment records, equipment conditions, quality assurance findings, and the responsibilities of appropriately qualified and authorized professionals.

International guidance provides valuable principles for managing these challenges. Viet Nam's updated atomic energy legislation and radiotherapy technical guidance establish an important domestic legal and professional context.

Together, these resources support an approach to resilience that considers the full treatment pathway rather than equipment availability alone.

A climate-ready radiotherapy service cannot guarantee that severe weather will never interrupt treatment. It can, however, become better at recognizing how interruptions affect patients, preserving reliable treatment information, and supporting appropriate clinical and technical decisions.

Ultimately, the goal is not simply to restart the radiation beam. It is to ensure that patients can continue receiving safe, accurate, and clinically appropriate cancer care.

Selected References and Policy Context
International Guidance and Evidence

International Atomic Energy Agency. Radiation Protection and Safety in Medical Uses of Ionizing Radiation. IAEA Specific Safety Guide No. SSG-46. 2018. Jointly sponsored by IAEA, ILO, PAHO, and WHO.

International Atomic Energy Agency. Radiation Protection and Safety of Radiation Sources: International Basic Safety Standards. IAEA General Safety Requirements Part 3. 2014.

American Association of Physicists in Medicine. Sutlief S, Buckey C, Ibbott G, et al. AAPM Task Group Report 314: Fault Recovery in External Beam Radiation Therapy. Medical Physics. 2025;52(1):21–44. Published online November 15, 2024. DOI: 10.1002/mp.17502.

The Royal College of Radiologists. Timely Delivery of Radical Radiotherapy: Guidelines for the Management of Unscheduled Treatment Interruptions. Fourth edition. 2019.

Müller-Polyzou R, Reuter-Oppermann M. Radiotherapy Continuity for Cancer Treatment: Lessons Learned From Natural Disasters. PLOS One. 2025;20(9). Published September 3, 2025. DOI: 10.1371/journal.pone.0308056.

Han-Oh S, Reuter-Oppermann M, Ozaki A, et al. Building Resilient Radiotherapy Systems: Lessons and Recommendations for Natural Disaster Risk Mitigation. Seminars in Radiation Oncology. 2026;39:151049. Published September 2026. DOI: 10.1016/j.semradonc.2026.151049.

Li H, Anderson R, Rong Y. Emergency Preparedness for External Beam Radiotherapy Machines: Treatment Continuity and Interruption Recovery in Radiation Oncology. Seminars in Radiation Oncology. 2026;39:151060. Published September 2026. DOI: 10.1016/j.semradonc.2026.151060.

Ford E. Resilience to Disaster: System Design and Future Directions. Seminars in Radiation Oncology. 2026;39:151055. Published September 2026. DOI: 10.1016/j.semradonc.2026.151055.

World Health Organization. Operational Framework for Building Climate Resilient and Low Carbon Health Systems. November 9, 2023.

World Health Organization. WHO Guidance for Climate-Resilient and Environmentally Sustainable Health Care Facilities. October 12, 2020.

Viet Nam Policy and Regulatory Context

National Assembly. Law on Atomic Energy No. 94/2025/QH15, dated June 27, 2025. Effective January 1, 2026, subject to applicable provisions and amendments.

Government of Viet Nam. Decree No. 332/2025/ND-CP, dated December 18, 2025, detailing provisions of the Law on Atomic Energy concerning radiation and nuclear safety, security, licensing, inspection, and incident response. Effective January 1, 2026.

Ministry of Science and Technology. Circular No. 59/2025/TT-BKHCN, dated December 31, 2025, regulating radiation safety and response to radiation and nuclear incidents. Effective January 1, 2026.

National Assembly. Law on Medical Examination and Treatment No. 15/2023/QH15, dated January 9, 2023. Effective January 1, 2024, as amended.

Government of Viet Nam. Decree No. 96/2023/ND-CP, dated December 30, 2023, detailing certain provisions of the Law on Medical Examination and Treatment, as amended where applicable.

Ministry of Health. Circular No. 23/2024/TT-BYT, dated October 18, 2024, promulgating the list of professional and technical procedures in medical examination and treatment, as amended where applicable.

Ministry of Health. Decision No. 1585/QD-BYT, dated June 3, 2026, promulgating Technical Procedure Guidelines for Radiotherapy, Volume 1, containing 34 technical procedures. Effective on the date of issuance.

Ministry of Health. Decision No. 1586/QD-BYT, dated June 3, 2026, promulgating Technical Procedure Guidelines for Radiotherapy, Volume 2, containing eight technical procedures.

Ministry of Health. Decision No. 1986/QD-BYT, dated July 1, 2026, amending the effective dates or application dates of selected technical guideline decisions. The application date of radiotherapy Volume 2 under Decision No. 1586/QD-BYT was changed to January 1, 2028.

Ministry of Health. Decision No. 3338/QD-BYT, dated September 9, 2013, promulgating oncology technical procedure guidelines, subject to the procedures specifically repealed by Decision No. 1585/QD-BYT and other applicable subsequent changes.

National Assembly. Law on Personal Data Protection No. 91/2025/QH15, effective January 1, 2026, where applicable to the processing and sharing of patient information.

Government of Viet Nam. Decree No. 356/2025/ND-CP, dated December 31, 2025, detailing and guiding the implementation of the Law on Personal Data Protection. Effective January 1, 2026.

Professional Scope and Important Notice

This article is intended for professional education, evidence synthesis, and general knowledge sharing. It does not constitute individualized cancer treatment advice, radiation treatment planning, radiobiological dose calculation, medical physics quality assurance instructions, equipment commissioning, radiation safety authorization, legal advice, or regulatory, accreditation, or certification requirements.

The scenarios and management considerations discussed are illustrative. They do not establish universal treatment interruption thresholds, clinical priorities, compensatory dose calculations, fractionation adjustments, quality assurance acceptance criteria, equipment release requirements, or conditions under which an individual radiotherapy treatment may resume.

Decisions concerning missed or partially delivered fractions, treatment modifications, replanning, equipment testing, radiation safety, patient transfer, and clinical resumption must be made by appropriately qualified and authorized professionals according to applicable Vietnamese laws and regulations, approved clinical procedures, relevant technical standards, manufacturer requirements, and the specific circumstances of the patient and healthcare facility.

International IAEA, AAPM, and professional guidance is presented for educational reference. Such guidance does not automatically constitute binding legal requirements in Viet Nam or authorize departures from applicable clinical and radiation safety obligations.

Climate & Health Vietnam does not claim authority to prescribe or modify radiotherapy, calculate compensatory radiation doses, validate treatment plans, certify equipment safety, authorize clinical treatment resumption, or determine the regulatory compliance of radiotherapy facilities.

References to IAEA, WHO, AAPM, professional colleges, and scientific publications do not imply their endorsement of CHVN or its activities.